Related Experiment Video
Updated: Mar 30, 2026

Insertion, Maintenance, and Removal of the Percutaneous Dual Lumen Cannula Right Ventricular Assist Device
Published on: July 20, 2022
Rectus-Sparing Technique for Driveline Insertion of Ventricular Assist Device
S Yukiko Asaki1, E Dean McKenzie2, Barbara Elias2
1Department of Cardiology, Texas Children's Hospital-Baylor College of Medicine, Houston, Texas.
Abstract:
We present a new technique for driveline insertion of the HeartWare ventricular assist device (HVAD) designed to preserve the integrity of the abdominal wall structure. Because of the size of the HVAD driveline connector (12 mm in diameter: triple the size of the driveline cable), the standard tunneling maneuver can result in tearing of the abdominal wall muscle layer, which is a primary mechanism to prevent ascending driveline infection. We find that our technique is particularly useful in children because their abdominal wall muscles are more fragile and thereby prone to accidental injury with blunt penetration when the standard technique is used.

