Monitoring the respiratory rate by miniature motion sensors in premature infants: a comparative study
Objective:
Existing respiratory rate (RR) monitors suffer from inaccuracy. The study assesses the accuracy of a novel modality that monitors lung ventilation with miniature motion sensors.
Study Design:
RR was measured by three methods: impedance technology, motion sensors and visual count, in babies (n=9) that breathed spontaneously or with respiratory support and babies (n=12) that received high-frequency oscillatory ventilation (HFOV).
Results:
A line close to equality (slope=0.96, r(2)=0.83) was obtained between the motion sensor and the visual count of the RR with narrow 95% limits of agreements (<14.0 b.p.m.). The relationship between the impedance and the visual count showed a lower correlation (r(2)=0.65) and wider 95% limits of agreements (21.4 b.p.m.). The motion sensor- and the ventilator-determined RRs demonstrated a good agreement during HFOV, whereas the impedance failed to measure the RR during HFOV.
Conclusion:
Monitoring RR with motion sensors is more accurate compared with the impedance, in infants, in all ventilation modes.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Respiratory Volumes and Capacities I
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...