Comparability and repeatability of pachymetry in keratoconus using four noncontact techniques.
Mukesh Kumar1, Rohit Shetty, Chaitra Jayadev
1Department of Cornea and Refractive Surgery, Narayana Nethralaya Super Specialty Eye Hospital, Bengaluru, Karnataka, India.
Central corneal thickness (CCT) measurements in keratoconus eyes showed good correlation between four devices. However, numerical differences suggest instruments are not interchangeable for clinical use.
Area of Science:
- Ophthalmology
- Corneal Imaging
- Keratoconus Research
Background:
- Accurate central corneal thickness (CCT) measurement is crucial for diagnosing and managing keratoconus.
- Noncontact pachymetry devices are widely used, but their agreement in keratoconus eyes requires evaluation.
Purpose of the Study:
- To compare the repeatability of CCT measurements using four noncontact pachymetry instruments.
- To assess interdevice agreement and correlation for CCT measurements in keratoconus patients.
Main Methods:
- CCT was measured in 50 eyes of 25 keratoconus patients using swept source OCT (SS-OCT), Pentacam (Scheimpflug), Orbscan IIz (scanning slit), and hand-held SD-OCT (HHSD-OCT).
- Test-retest variability, measurement correlation, and interdevice agreement were analyzed.
Main Results:
- All four devices demonstrated good correlation (r > 0.9, P < 0.001) in CCT measurements.
- Mean CCT values ranged from 447 ± 42 mm (Pentacam) to 462 ± 41 mm (HHSD-OCT).
- HHSD-OCT overestimated CCT, and significant numerical differences were observed between all device pairs, despite good correlation.
Conclusions:
- While CCT measurements from the four devices correlate well, the numerical discrepancies limit their interchangeable use in clinical practice.
- Further research is needed to establish the clinical utility of each device for specific keratoconus management decisions.