Practical dose delivery verification of craniospinal IMRT.
Young K Lee1, Anthony T Kim, Peiying Zhao
1Odette Cancer Centre, Sunnybrook Health Sciences Centre and University of Toronto. young.lee@sunnybrook.ca.
Journal of Applied Clinical Medical Physics
|December 25, 2015
Summary
This study validates dosimetry methods for craniospinal intensity-modulated radiation therapy (IMRT) verification. The findings confirm that ionization chambers, MapCHECK, and ArcCHECK can accurately verify complex IMRT dose delivery for patients.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Physics
Background:
- Intensity-modulated radiation therapy (IMRT) for craniospinal irradiation (CSI) offers improved organ-at-risk (OAR) sparing.
- The complexity of multi-isocenter CSI necessitates robust pre-treatment dose verification methods.
- Accurate dose delivery verification is crucial for patient safety and treatment efficacy in CSI.
Purpose of the Study:
- To propose and evaluate dosimetry techniques for commissioning and patient-specific verification of craniospinal IMRT.
- To compare the accuracy and clinical applicability of ionization chambers, MapCHECK, and ArcCHECK for CSI dose verification.
- To establish acceptable criteria for dose difference and gamma index in the verification of craniospinal IMRT.
Main Methods:
- Retrospective planning of 10 conventional CSI patients using a 3-isocenter IMRT technique.
- Dose verification using ionization chambers in junction regions, MapCHECK in overlap regions, and ArcCHECK for all beams.
- Defined clinical acceptability criteria: ±3% dose difference for ionization chambers, and ≥95% of points with a γ-index < 1 for 3%/3 mm for MapCHECK and ArcCHECK.
Main Results:
- Median planned to measured dose differences were within ±3% for ionization chambers.
- Median percentage of points passing the γ-index < 1 (3%/3 mm) criterion was 98.0% for MapCHECK and 99.7% for ArcCHECK.
- All tested methods demonstrated clinically acceptable agreement between planned and measured doses.
Conclusions:
- Ionization chambers, MapCHECK, and ArcCHECK are suitable for verifying craniospinal IMRT dose delivery.
- These methods can be employed for both commissioning the treatment technique and for patient-specific pre-treatment verification.
- The combined use of these techniques provides comprehensive quality assurance for craniospinal IMRT.