Related Experiment Video
Updated: Mar 28, 2026

Double-barreled and Concentric Microelectrodes for Measurement of Extracellular Ion Signals in Brain Tissue
Published on: September 5, 2015
Phenytoin speciation with potentiometric and chronopotentiometric ion-selective membrane electrodes
Sutida Jansod1, Majid Ghahraman Afshar1, Gastón A Crespo1
1Department of Inorganic and Analytical Chemistry, University of Geneva, Quai Ernest-Ansermet 30, CH-1211 Geneva, Switzerland.
Abstract:
We report on an electrochemical protocol based on perm-selective membranes to provide valuable information about the speciation of ionizable drugs, with phenytoin as a model example. Membranes containing varying amounts of tetradodecylammonium chloride (TDDA) were read out at zero current (potentiometry) and with applied current techniques (chronopotentiometry). Potentiometry allows one to assess the ionized form of phenytoin (pKa~8.2) that corresponds to a negatively monocharged ion. A careful optimization of the membrane components resulted in a lower limit of detection (~1.6 µM) than previous reports. Once the pH (from 9 to 10) or the concentration of albumin is varied in the sample (from 0 to 30 g L(-1)), the potentiometric signal changes abruptly as a result of reducing/increasing the ionized concentration of phenytoin. Therefore, potentiometry as a single technique is by itself not sufficient to obtain information about the concentration and speciation of the drug in the system. For this reason, a tandem configuration with chronopotentiometry as additional readout principle was used to determine the total and ionized concentration of phenytoin. In samples containing excess albumin the rate-limiting step for the chronopotentiometry readout appears to be the diffusion of ionized phenytoin preceded by comparatively rapid deprotonation and decomplexation reactions. This protocol was applied to measure phenytoin in pharmaceutical tables (100mg per tablet). This tandem approach can likely be extended to more ionizable drugs and may eventually be utilized in view of pharmacological monitoring of drugs during the delivery process.
Related Concept Videos
Potentiometry: Membrane Electrodes
Potentiometry: Overview
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Potentiometric Titration: Overview
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in...
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential...