DOSE REFERENCE LEVELS IN SPANISH INTRAORAL DENTAL RADIOLOGY: STABILISATION OF THE INCORPORATION OF DIGITAL SYSTEMS IN
M Alcaraz1, F Velasco2, A Olivares2
1Department of Radiology and Physical Medicine, Faculty of Medicine/Dentistry, University of Murcia, Campus Espinardo, Murcia 30100, Spain mab@um.es.
Abstract:
A total of 34 044 official quality assurance reports in dental radiodiagnostic surgery from 16 regions of Spain, compiled from 2002 to 2014, were studied in order to determine the progress of diagnostic reference levels (DRLs) for obtaining diagnostic images under normal conditions for clinical practice in Spanish dental clinics. A DRL of 2.8 mGy was set in 2014, which represents a 41.7 % decrease compared with that of 2002 (4.8 mGy). Over the same time period, the mean dose fell by 55.2 %. However, over the last 3 y, the stabilisation of the mean dose administered to patients has been observed with only a 6.7 % reduction in DRLs, which corresponds to the stabilisation of dental radiodiagnostic surgery on replacing the use of radiographic film with digital imaging systems.
Related Concept Videos
Dosage Regimen: Multiple Oral Dosage
X-ray Imaging
Biological Effects of Radiation
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant
Radiological Investigation I: X-ray and CT
Dosage Regimen: Fixed Dose
Fixed-dose regimens can be used for various routes of administration, including intravenous (IV) injections and oral medications. For IV administration, a predetermined amount of the drug is...