Effect of mA Reduction on Image Quality Parameters and Patient Dose in Computed Tomography Imaging
Purpose:
To establish the extent of dose reduction that can be achieved using a computed tomography (CT) quality assurance phantom with various user-defined protocols at different milliamperage (mA) settings and kilovoltage peak (kVp) without affecting image quality.
Methods:
Images of a CT quality assurance phantom containing line-pair patterns and 3 tissue-equivalent inserts in water were acquired with different mA settings on 4 different CT scanners. The kVp was kept constant at 120 while the mA was changed in increments of 20, from 30 mA to 190 mA. Three images from each series acquired at different mA settings were selected: one each for the quantitative analysis of spatial resolution, contrast-to-noise ratio (CNR), and image noise using image analysis freeware.
Results:
Statistical analysis was carried out with a statistical probability of P < .05 indicating significant difference. Results of the statistical analysis revealed significant differences in the CNR of all 3 types of tissue-equivalent inserts (P < .001) and in image noise (P < .001) as the mA was changed. However, changes in mA did not produce significant differences in the spatial resolution of the images (P > .06).
Conclusion:
A radiation dose reduction of 25% was achieved in 4 different CT scanners without significantly affecting the 3 image quality parameters.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography
Magnetic Resonance Imaging
X-ray Imaging
Radiological Investigation I: X-ray and CT