Related Experiment Video
Updated: Mar 28, 2026

Using Synchrotron Radiation Microtomography to Investigate Multi-scale Three-dimensional Microelectronic Packages
Published on: April 13, 2016
Monte-Carlo simulation of a slot-scanning X-ray imaging system
Mayuresh Kulkarni1, Ronald Dendere1, Fred Nicolls2
1Division of Biomedical Engineering, University of Cape Town, Cape Town, Western Cape, South Africa.
Abstract:
We present a method for simulating slot-scanning X-ray imaging using the general-purpose Monte Carlo simulation package PENELOPE and penEasy Imaging. Different phantoms can be defined with the PENGEOM package, which defines bodies as combinations of volumes limited by quadric surfaces. The source-detector geometry, the position of the object, the collimator, the X-ray tube properties, the detector material and the pixel dimensions are defined. The output of the time-delay integration detector is simulated using sequential slot outputs derived from penEasy Imaging. The simulations are validated using tungsten and aluminium test objects, which are both simulated and imaged. The simulations are compared to the X-ray images using standard image quality metrics. The MTF, NPS and DQE curves show that the real and simulated X-ray images are comparable in terms of spatial resolution, noise and frequency information. The implementation can be modified to suit alterations in the system being simulated.