A time delay correction technique for SS-OCT
Background:
K-trigger signals from light source or external Mach-Zehnder interferometer are commonly used in swept source optical coherence tomography (SS-OCT) to achieve interference signal sampling and resolution enhancement. But extra signal transmission time is required due to the increased complexity of interference system using k-trigger signals, resulting in a misalignment (a.k.a., time delay) between the k-trigger and the interference signals.
Objective:
The objective of this paper is to propose a numerical technique to correct the time delay between these two signals.
Methods:
The time delay is iterated in this technique and each iterated correction result is evaluated through calculating the full width at half maximum (FWHM) of peaks in time-domain. The precise alignment between the k-trigger and interference signals coincides with the minimum value of FWHM.
Results:
The experimental results show that the FWHM can be improved by up to 50% using this time delay correction technique.
Conclusions:
The proposed technique assists to provide precise sample tomography, thus enhancing the way of understanding sample characteristic.