Related Experiment Video
Updated: Mar 27, 2026

Fully Endoscopic Mitral Valve Repair with Percutaneous Cannulation of Groin Vessels
Published on: May 26, 2023
Rapid Switch from Intra-Aortic Balloon Pumping to Percutaneous Cardiopulmonary Support Using Perclose ProGlide
Kenichi Sakakura1, Yusuke Adachi1, Yousuke Taniguchi1
1Division of Cardiovascular Medicine, Saitama Medical Center, Jichi Medical University, 1-847 Amanuma, Omiya, Saitama 330-8503, Japan.
Abstract:
We present a case of a patient who needed rapid switch from intra-aortic balloon pumping (IABP) to percutaneous cardiopulmonary support (PCPS)/venoarterial extracorporeal membrane oxygenation. It is difficult to switch from IABP to PCPS, because 0.035-inch guidewires cannot pass the IABP guidewire lumen (0.025-inch compatible), and the IABP sheath needs to be removed together with the IABP catheter. First, a 0.025-inch guidewire was inserted into the IABP wire lumen, and then the IABP catheter together with the 8 Fr IABP sheath was removed, leaving the 0.025-inch guidewire in place. We used the Perclose ProGlide for safe and rapid exchange of the 0.025-inch guidewire for a 0.035-inch guidewire. This allowed insertion of a PCPS cannula and the prompt initiation of PCPS.

