Related Experiment Video
Updated: Mar 26, 2026

Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry UPLC-HRMS
Published on: May 20, 2013
Fully automated semi-quantitative toxicological screening in three biological matrices using turbulent flow
Lucie Roche1, Jérémy Pinguet1, Pauline Herviou1
1CHU Clermont-Ferrand, Service de Pharmacologie Médicale, Laboratoire de pharmacologie-toxicologie, F-63003 Clermont-Ferrand, France; Inserm, UMR 1107 Neuro-Dol, F-63001 Clermont-Ferrand, France.
Background:
In clinical and forensic toxicology, fast and specific methods are needed for the screening of different classes of drugs. A complete general unknown screening procedure was developed using turbulent flow chromatography with electrospray ionization and Orbitrap mass spectrometry.
Methods:
After protein precipitation, samples were injected directly into the turbulent flow chromatographic system and analyzed with an Orbitrap mass spectrometer. The Exactive® operated in positive and negative modes with alternated high collision dissociation in order to obtain characteristic fragments. We built a library containing 616 compounds by analyzing a reference standard for all the molecules.
Results:
Identification was based on retention time, accurate measured mass, isotopic pattern and presence of specific fragments. For each substance, we set a calibration range encompassing infra-therapeutic, therapeutic, supra-therapeutic and toxic concentrations in order to generate semi-quantitative result. For 65% of the components, the limit of detection was below 5 ng/mL. The validation process showed the approach to be selective, sensitive, accurate and precise.
Conclusion:
The method has been accredited by COFRAC (French Accreditation Committee) according to the ISO 15189 standard. Applicability was successfully tested by analyzing authentic serum, urine and whole blood samples.