Related Experiment Video
Updated: Mar 26, 2026

Synthesis of Indoxyl-glycosides for Detection of Glycosidase Activities
Published on: May 27, 2015
Development of Silicalite/Glucose Oxidase-Based Biosensor and Its Application for Glucose Determination in Juices and
Oleksandr Ye Dudchenko1,2, Viktoriya M Pyeshkova3,4, Oleksandr O Soldatkin5,6
1Laboratory of Biomolecular Electronics, Institute of Molecular Biology and Genetics, National Academy of Sciences of Ukraine, 150 Zabolotnogo St., 03143, Kyiv, Ukraine. dc182@yandex.ru.
Abstract:
The application of silicalite for improvement of enzyme adsorption on new stainless steel electrodes is reported. Glucose oxidase (GOx) was immobilized by two methods: cross-linking by glutaraldehyde (GOx-GA) and cross-linking by glutaraldehyde along with GOx adsorption on silicalite-modified electrode (SME) (GOx-SME-GA). The GOx-SME-GA biosensors were characterized by a four- to fivefold higher sensitivity than GOx-GA biosensor. It was concluded that silicalite together with GA sufficiently enhances enzyme adhesion on stainless steel electrodes. The developed GOx-SME-GA biosensors were characterized by good reproducibility of biosensor preparation (relative standard deviation (RSD)-18 %), improved signal reproducibility (RSD of glucose determination was 7 %), and good storage stability (29 % loss of activity after 18-day storage). A series of fruit juices and nectars was analyzed using GOx-SME-GA biosensor for determination of glucose concentration. The obtained results showed good correlation with the data of high-performance liquid chromatography (HPLC) (R = 0.99).