Repeatability Using Automatic Tracing with Canon OCT- HS100 and Zeiss Cirrus HD-OCT 5000
Rune Brautaset1, Ulrika Birkeldh1, Petra Frehr Alstig1
1Unit of Optometry, Department of Clinical Neuroscience, St. Erik Eye Hospital, Karolinska Institutet, Stockholm, Sweden.
Automatic tracking in optical coherence tomography (OCT) enhances measurement repeatability for retinal imaging. However, data from Canon and Zeiss OCT devices remain non-interchangeable.
Area of Science:
- Ophthalmology
- Medical Imaging
- Biomedical Engineering
Background:
- Optical coherence tomography (OCT) provides high-resolution retinal images crucial for diagnosing and monitoring eye diseases.
- Accurate and repeatable measurements are essential for clinical applications of OCT in ophthalmology.
- Assessing the repeatability of OCT devices is vital for reliable patient care.
Purpose of the Study:
- To evaluate the repeatability of macular and optic disc measurements using Canon OCT-HS100 and Zeiss Cirrus HD-OCT 5000.
- To compare the repeatability of these two OCT instruments.
- To determine the impact of automatic tracking on measurement repeatability.
Main Methods:
- Repeatability analysis of macula and optic disc parameters from 30 healthy subjects using Canon and Zeiss OCT devices.
- Evaluation of repeatability using coefficient of repeatability (CR) and CR percentage of the mean (CR%).
- Comparison of measurements between the two OCT instruments.
Main Results:
- Coefficient of repeatability percentages (CR%) for optic disc parameters ranged from 0.90-22.22% (Canon) and 0.00-16.00% (Zeiss).
- CR% for macular parameters ranged from 0.62-2.81% (Canon) and 0.99-1.81% (Zeiss).
- Repeatability significantly improved with automatic tracking systems compared to previous data for both devices.
Conclusions:
- The automatic tracking function significantly improves measurement repeatability in both Canon and Zeiss OCT devices.
- Despite improved repeatability, measurements obtained from the Canon OCT-HS100 and Zeiss Cirrus HD-OCT 5000 are not interchangeable.
- Further research may be needed to understand the clinical implications of non-interchangeable OCT measurements.