Torsional power study using CENTURION phacoemulsification technology
Jason D Jensen1, Dallas S Shi1, Mark S Robinson1
1Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, Salt Lake City, Utah, USA.
Background:
To evaluate the effect of varying levels of power on phacoemulsification efficiency using the CENTURION Vision System.
Methods:
Formalin-soaked porcine lenses were divided into 2-mm cubes; 0.9-mm, balanced tips were used. Torsional power levels were tested from 10% to 100% at 10% intervals. Vacuum was set to 550 mmHg, aspiration to 50 ml/min, and intraocular pressure at 50 mmHg. Efficiency (time to lens removal) and chatter (number of lens fragment repulsions from the tip) were determined.
Results:
Increasing torsional power up to 60% increased efficiency. This effect was linear from 30 to 60% power (R2 = .90; P < 0.05). There were no significant differences in efficiency past 60%. Chatter was highest at 10% power and decreased linearly (R2 = .87; P = 0.007) as power was increased up to 60% power, and chatter did not improve above this power level.
Conclusions:
Power improved efficiency only up to a 60% power level, and then was negligible. Chatter correlated well with power up to the 60% level, so that as power was increased, chatter decreased. Because there are no additional benefits in efficiency past 60% power, and because chatter is minimal at 60% power, we recommend torsional ultrasound at 60% as the optimal power setting for using the CENTURION System for phacoemulsification.l.
Related Concept Videos
Mechanism of Ciliary Motion
The cilia are made up of microtubules in a 9+2 arrangement, with nine microtubule doublet ring bundles, surrounding a pair of central singlet microtubule bundles. The doublet microtubule bundles are...
Centrifugal Force
Angle Closure Glaucoma: Treatment