Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

The use of anthropomorphic phantoms for patient dose and image contrast surveys in digital radiography.

Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine·2025
Same author

PATIENT AND STAFF DOSES FOR VARIOUS INTERVENTIONAL RADIOLOGY AND CARDIOLOGY EXAMINATIONS IN TURKEY.

Radiation protection dosimetry·2022
Same author

Correlation between physical measurements and observer evaluations of image quality in digital chest radiography.

Medical physics·2020
Same author

Comparison of noise reduction methods in photoacoustic microscopy.

Computers in biology and medicine·2019
Same author

Investigating the effect of dental implant materials with different densities on radiotherapy dose distribution using Monte-Carlo simulation and pencil beam convolution algorithm.

Dento maxillo facial radiology·2019
Same author

Investigation of noise sources for digital radiography systems.

Radiological physics and technology·2016

Related Experiment Video

Updated: Mar 23, 2026

Automated Quantification and Analysis of Cell Counting Procedures Using ImageJ Plugins
11:01

Automated Quantification and Analysis of Cell Counting Procedures Using ImageJ Plugins

Published on: November 17, 2016

49.4K

Investigation of grid performance using simple image quality tests.

Dogan Bor1, Ozlem Birgul2, Umran Onal2

  • 1Department of Physics Engineering, Faculty of Engineering, Ankara University, 06100 Ankara, Turkey.

Journal of Medical Physics
|April 7, 2016
PubMed
Summary

Antiscatter grids enhance X-ray image contrast but increase radiation dose. This study links grid performance parameters to image quality metrics, offering a practical method for assessing grid effectiveness in digital radiography.

Keywords:
Bucky factorcontrastgrid performanceimage quality

More Related Videos

Cryo-Electron Microscopic Grid Preparation for Time-Resolved Studies using a Novel Robotic System, Spotiton
08:59

Cryo-Electron Microscopic Grid Preparation for Time-Resolved Studies using a Novel Robotic System, Spotiton

Published on: February 25, 2021

4.4K
Quantifying Microglia Morphology from Photomicrographs of Immunohistochemistry Prepared Tissue Using ImageJ
08:44

Quantifying Microglia Morphology from Photomicrographs of Immunohistochemistry Prepared Tissue Using ImageJ

Published on: June 5, 2018

73.6K

Related Experiment Videos

Last Updated: Mar 23, 2026

Automated Quantification and Analysis of Cell Counting Procedures Using ImageJ Plugins
11:01

Automated Quantification and Analysis of Cell Counting Procedures Using ImageJ Plugins

Published on: November 17, 2016

49.4K
Cryo-Electron Microscopic Grid Preparation for Time-Resolved Studies using a Novel Robotic System, Spotiton
08:59

Cryo-Electron Microscopic Grid Preparation for Time-Resolved Studies using a Novel Robotic System, Spotiton

Published on: February 25, 2021

4.4K
Quantifying Microglia Morphology from Photomicrographs of Immunohistochemistry Prepared Tissue Using ImageJ
08:44

Quantifying Microglia Morphology from Photomicrographs of Immunohistochemistry Prepared Tissue Using ImageJ

Published on: June 5, 2018

73.6K

Area of Science:

  • Medical Physics
  • Radiological Imaging
  • Diagnostic Radiology

Background:

  • Antiscatter grids are crucial for improving X-ray image contrast in digital radiography.
  • However, their use increases patient radiation dose, necessitating careful selection.
  • Understanding grid performance is vital, particularly in pediatric digital imaging.

Purpose of the Study:

  • To investigate the relationship between antiscatter grid performance parameters and numerical image quality metrics.
  • To establish a practical method for evaluating grid performance in digital radiological examinations.

Main Methods:

  • Grid parameters (bucky factor, selectivity, contrast improvement factor, signal-to-noise improvement factor) were measured using a digital fluoroscopic system.
  • Measurements were conducted across various polymethyl methacrylate phantom thicknesses (5-25 cm) and tube voltages (70-120 kVp).
  • Image contrast and spatial resolution were assessed using phantoms, and bucky factor and signal-to-noise improvement factor were calculated from images with and without grids.

Main Results:

  • A strong correlation was observed between bucky factor values derived from grid parameters and image quality metrics.
  • The study found good agreement between the two methods of assessing bucky factor.
  • The proposed approach demonstrated a practical way to estimate grid performance.

Conclusions:

  • A validated, practical method exists for estimating antiscatter grid performance in digital fluoroscopic examinations.
  • This approach aids in optimizing grid selection to balance image quality and radiation dose.
  • Further application in pediatric digital imaging is recommended.