The Rigid Tube as an Alternative in Controlling the Problematic Airway
Published on: June 6, 2020
Use-Dependent Curvature Changes in the GlideRite® Reusable Intubation Stylet
Cameron R Smith1, Felipe Urdaneta, Nikolaus Gravenstein
1From the *Department of Anesthesiology, University of Florida College of Medicine, Gainesville, Florida; and †Department of Anesthesiology, North Florida/South Georgia Veteran's Health System, Gainesville, Florida.
Reusable GlideRite stylets used with Glidescope video laryngoscopes lose their intended curvature with repeated use and sterilization. This shape change may impact intubation difficulty.
Area of Science:
- Medical Devices
- Anesthesiology
- Surgical Technology
Background:
- The Glidescope® is a widely used video laryngoscope.
- It is commonly paired with the reusable GlideRite® stylet, designed with a specific curvature.
- The long-term shape retention of these 'nonmalleable' stylets after repeated use and sterilization is not well understood.
Purpose of the Study:
- To investigate whether GlideRite® stylets maintain their original shape and curvature through repeated clinical use and sterilization.
- To assess the impact of repeated insertion and removal on stylet shape.
Main Methods:
- High-resolution digital photography of 55 GlideRite® stylets (5 new, 50 from operating room stock) against a grid background.
- Analysis of stylet curvature using Adobe Photoshop®.
- A new stylet underwent 100 insertion/removal cycles from an endotracheal tube, with photography every 20 cycles to track shape changes.
Main Results:
- New GlideRite® stylets exhibited a consistent handle-to-tip angle (23.44° ± 1.04°).
- Stylets from clinical use showed significant variations in curvature, with some straighter (18.39° ± 2.59°) and others more curved (27.65° ± 2.73°) than new ones.
- Repeated insertion/removal cycles caused a new stylet to gradually straighten from 22° to 19.2°.
Conclusions:
- Used GlideRite® reusable stylets do not reliably maintain their original shape or curvature compared to new ones.
- Repeated clinical use likely causes stylet straightening, similar to laboratory findings.
- Practitioners may manually bend stylets, potentially altering the intended curvature and impacting intubation effectiveness.
More Related Videos
Related Concept Videos
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Suctioning the Nasopharyngeal Airway
Equipment Required