Related Experiment Video
Updated: Mar 22, 2026

Microdialysis of Ethanol During Operant Ethanol Self-administration and Ethanol Determination by Gas Chromatography
Published on: September 5, 2012
[Determination of ethylene glycol in workplace air by capillary column gas chromatography]
Objective:
To establish the method of capillary column gas chromatography for determination of ethylene glycol in workplace air.
Methods:
Ethylene glycol in workplace air was collected with silicone tube, desorbed with methanol, separated with FFAP (nitroterephthalic acid-modified polyethylene glycol)capillary column, and measured with flame ionization detector.
Results:
The detection limit of ethylene glycol was 0.41 mg/L, the lower limit of quantification was 1.4 mg/L, the range of measurement was 1.4~163.9 mg/L, and the minimum detectable concentration was 0.3 mg/m3 (1.5 L of air was collected as the sample). This method had a good repeatability, the relative standard deviation was 1.4%~5.2%, the average desorption efficiency was 94.4%~101.7%, and the sampling efficiency was 99.2%~100%. The penetrating capacity of 200 mg silicone was higher than 6.9 mg, and the samples could be preserved for 14 days at room temperature.
Conclusion:
The method has a low detection limit, high accuracy, and good precision, which is feasible for determination of ethylene glycol in workplace air.
More Related Videos
08:33Use of Capillary Aerosol Generator in Continuous Production of Controlled Aerosol for Non-Clinical Studies
Published on: April 12, 2022
11:06GC-based Detection of Aldononitrile Acetate Derivatized Glucosamine and Muramic Acid for Microbial Residue Determination in Soil
Published on: May 19, 2012
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....