Individual selection of X-ray tube settings in computed tomography coronary angiography: Reliability of an automated
Tahir Durmus1, Reny Luhur1, Tareef Daqqaq1
1Department of Radiology, Charité-Universitätsmedizin Berlin, Charité Campus Mitte, Berlin, Germany.
Objectives:
To evaluate a software tool that claims to maintain a constant contrast-to-noise ratio (CNR) in high-pitch dual-source computed tomography coronary angiography (CTCA) by automatically selecting both X-ray tube voltage and current.
Methods:
A total of 302 patients (171 males; age 61±12years; body weight 82±17kg, body mass index 27.3±4.6kg/cm(2)) underwent CTCA with a topogram-based, automatic selection of both tube voltage and current using dedicated software with quality reference values of 100kV and 250mAs/rotation (i.e., standard values for an average adult weighing 75kg) and an injected iodine load of 222mg/kg.
Results:
The average radiation dose was estimated to be 1.02±0.64mSv. All data sets had adequate contrast enhancement. Average CNR in the aortic root, left ventricle, and left and right coronary artery was 15.7±4.5, 8.3±2.9, 16.1±4.3 and 15.3±3.9 respectively. Individual CNR values were independent of patients' body size and radiation dose. However, individual CNR values may vary considerably between subjects as reflected by interquartile ranges of 12.6-18.6, 6.2-9.9, 12.8-18.9 and 12.5-17.9 respectively. Moreover, average CNR values were significantly lower in males than females (15.1±4.1 vs. 16.6±11.7 and 7.9±2.7 vs. 8.9±3.0, 15.5±3.9 vs. 16.9±4.6 and 14.7±3.6 vs. 16.0±4.1 respectively).
Conclusion:
A topogram-based automatic selection of X-ray tube settings in CTCA provides diagnostic image quality independent of patients' body size. Nevertheless, considerable variation of individual CNR values between patients and significant differences of CNR values between males and females occur which questions the reliability of this approach.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
06:57Author Spotlight: Advancing Cardiovascular Imaging - Introducing the Spatially Weighted Calcium Score for Early Disease Detection
Published on: September 22, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...