A Novel Free Flap Monitoring System Using Tissue Oximetry with Text Message Alerts
Joseph A Ricci1, Christina R Vargas1, Samuel J Lin1
1Division of Plastic and Reconstructive Surgery, Beth Israel Deaconess Medical Center, Boston, Massachusetts.
Abstract:
Background The time to detection of vascular compromise is a significant predictor of free flap salvage outcomes as early reexploration improves salvage rates. Continuous transcutaneous near-infrared tissue oximetry is an objective, quantitative method of detecting flap vascular compromise and has been shown to allow earlier reexploration and higher salvage rates than clinical assessment alone. We designed a novel text messaging system to improve communication using tissue oximetry monitoring. Methods A retrospective review was performed of a prospectively collected database of all microsurgical breast reconstructions from 2008 to 2015. A novel text messaging system was introduced in 2013 and programmed to send text messages alert when the tissue oximetry readings suggested potential flap compromise based on established thresholds. Patient demographics and complications, including rate of reexploration and flap loss were assessed. Results There were 900 autologous microsurgical breast free flaps during the study period: 614 were monitored with standard clinical monitoring and tissue oximetry compared with 286 flaps with the additional text messaging system. There were 27 unplanned returns to the operating room in the tissue oximetry group and 5 in the text messaging group with 1 complete flap loss in each group. Reexploration occurred sooner as a result of these text message alerts (17.5 vs. 26.6 hours postoperatively), however, it did not achieve statistical significance. Conclusions We were able to demonstrate the use of a novel text messaging system for tissue oximetry. This alert system shows promise in identifying impending flap loss with rapid notification of the surgical team.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Holter Monitor: 24-Hour Monitoring
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: