Oxygen cylinders are heavy and present a number of hazards, and liquid oxygen is too heavy and cumbersome to be used in far forward environments. Portable oxygen concentrators (POCs) and chemical oxygen generators (COGs) have been proposed as a solution. We evaluated 3 commercially available POCs and 3 COGs in a laboratory setting. Altitude testing was done at sea level and 8,000, 16,000, and 22,000 ft. Temperature extreme testing was performed after storing devices at 60°C and -35°C for 24 hours. Mean FIO2 decreased after storage at -35°C with Eclipse and iGo POCs and also at the higher volumes after storage at 60°C with the Eclipse. The iGo ceased to operate at 16,000 ft, but the Eclipse and Saros were unaffected by altitude. Oxygen flow, duration of operation, and total oxygen volume varied between COGs and within the same device type. Output decreased after storage at -35°C, but increased at each altitude as compared to sea level. This study showed significant differences in the performance of POCs and COGs after storage at temperature extremes and with the COGs at altitude. Clinicians must understand the performance characteristics of devices in all potential environments.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Oxygen Requirements and Growth Patterns
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue,...
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...