Related Experiment Video
Updated: Mar 20, 2026

Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
A 16-F Sheath with Endobronchial Forceps Improves Reported Retrieval Success of Long-Dwelling "Closed Cell" Inferior
Jeffrey Forris Beecham Chick1, S William Stavropoulos1, Benjamin J Shin1
1Division of Interventional Radiology, Department of Radiology, Hospital of the University of Pennsylvania, Perelman School of Medicine, 1 Silverstein, 3400 Spruce Street, Philadelphia, PA 19104.
Purpose:
To report results of a 16-F sheath/endobronchial forceps combination in retrieval of "closed cell" inferior vena cava (IVC) filter designs with prolonged dwell times of ≥ 60 days.
Materials And Methods:
Forceps retrieval of selected IVC filters using a 16-F 45-cm sheath was performed in 35 patients (25 women and 10 men). Patients presented with filters that were no longer needed without (n = 10) or with (n = 21) prior failed retrieval, caval thrombosis before lysis (n = 3), and back pain (n = 1). Filter designs included 19 Günther Tulip (54%), 10 Option (29%), and six OptEase (17%). Mean dwell time was 977 days. Imaging characteristics, technical success, fluoroscopy time, and complications were recorded.
Results:
Filters included 20 tip-embedded, 6 tip-adherent, 4 fractured, and 6 containing thrombus. All were strut and wall-embedded. Initial snare failures occurred in 8 filters because of hook straightening (n = 4), adherent material (n = 2), snare breakage (n = 1), and filter tilt (n = 1). The 16-F sheath/forceps combination was 100% successful in removing filters. In 1 patient, 2 extravascular fragments could not be retrieved. Median fluoroscopy time, excluding 3 extended lysis procedures, was 8 minutes. Minor caval abnormalities, including caval spasm and caval defects, were noted in 17 patients, and filter fracture occurred in 2 patients. No major complications occurred.
Conclusions:
A high rate of retrieval for closed cell long-dwelling strut and wall-embedded filters may be achieved using a 16-F sheath/endobronchial forceps combination.
Related Concept Videos
Venous Thrombosis III: Interprofessional Care
Hemodialysis I: Introduction
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...