A Monte Carlo study on (223)Ra imaging for unsealed radionuclide therapy
Akihiko Takahashi1, Kenta Miwa1, Masayuki Sasaki1
1Faculty of Medical Sciences, Department of Health Sciences, Kyushu University, 3-1-1, Maidashi, Higashi-ku, Fukuoka 812-8582, Japan.
Purpose:
Radium-223 ((223)Ra), an α-emitting radionuclide, is used in unsealed radionuclide therapy for metastatic bone tumors. The demand for qualitative (223)Ra imaging is growing to optimize dosimetry. The authors simulated (223)Ra imaging using an in-house Monte Carlo simulation code and investigated the feasibility and utility of (223)Ra imaging.
Methods:
The Monte Carlo code comprises two modules, hexagon and nai. The hexagon code simulates the photon and electron interactions in the tissues and collimator, and the nai code simulates the response of the NaI detector system. A 3D numeric phantom created using computed tomography images of a chest phantom was installed in the hexagon code. (223)Ra accumulated in a part of the spine, and three x-rays and 19 γ rays between 80 and 450 keV were selected as the emitted photons. To evaluate the quality of the (223)Ra imaging, the authors also simulated technetium-99m ((99m)Tc) imaging under the same conditions and compared the results.
Results:
The sensitivities of the three photopeaks were 147 counts per unit of source activity (cps MBq(-1); photopeak: 84 keV, full width of energy window: 20%), 166 cps MBq(-1) (154 keV, 15%), and 158 cps MBq(-1) (270 keV, 10%) for a low-energy general-purpose (LEGP) collimator, and those for the medium-energy general-purpose (MEGP) collimator were 33, 13, and 8.0 cps MBq(-1), respectively. In the case of (99m)Tc, the sensitivity was 55 cps MBq(-1) (141 keV, 20%) for LEGP and 52 cps MBq(-1) for MEGP. The fractions of unscattered photons of the total photons reflecting the image quality were 0.09 (84 keV), 0.03 (154 keV), and 0.02 (270 keV) for the LEGP collimator and 0.41, 0.25, and 0.50 for the MEGP collimator, respectively. Conversely, this fraction was approximately 0.65 for the simulated (99m)Tc imaging. The sensitivity with the LEGP collimator appeared very high. However, almost all of the counts were because of photons that penetrated or were scattered in the collimator; therefore, the proportions of unscattered photons were small.
Conclusions:
Their simulation study revealed that the most promising scheme for (223)Ra imaging is an 84-keV window using an MEGP collimator. The sensitivity of the photopeaks above 100 keV is too low for (223)Ra imaging. A comparison of the fractions of unscattered photons reveals that the sensitivity and image quality are approximately two-thirds of those for (99m)Tc imaging.
More Related Videos
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...