Related Experiment Video
Updated: Mar 19, 2026

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
A model of CT dose profiles in Banach space; with applications to CT dosimetry
1Baylor Scott & White Health, Dallas, TX 75246, USA.
Abstract:
In this paper the scatter component of computed tomography dose profiles is modeled using the solution to a nonlinear ordinary differential equation. This scatter function is summed with a modeled primary function of approximate trapezoidal shape. The primary dose profile is modeled to include the analytic continuation of the Heaviside step function. A mathematical theory is developed in a Banach space. The modeled function is used to accurately fit data from a 256-slice GE Revolution scanner. A 60 cm long body phantom is assembled and used for data collection with both a pencil chamber and a Farmer-type chamber.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Dose Response Curve: Conventional Versus Nonmonotonic
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Dosage Regimen: Multiple Oral Dosage
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Dosage Regimens: Partial Pharmacokinetic Parameters