Related Experiment Video
Updated: Mar 19, 2026

Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
Published on: January 8, 2019
Tidal volume measurement error in pressure control modes of mechanical ventilation: A model study
Robert L Chatburn1, Eduardo Mireles-Cabodevila1, Madhu Sasidhar1
1Respiratory Institute, Cleveland Clinic, Mail stop M56, 9500 Euclid Ave, Cleveland, OH, United States.
Unlabelled:
Tidal volume (VT) measurement during pressure control (PC) ventilation with preset inspiratory time may produce errors due to patient inspiratory effort. We evaluated VT error in 3 common ICU ventilators.
Methods:
Simulated patient: 60kg adult with ARDS using IngMar Medical ASL 5000 having moderate inspiratory effort. Ventilators evaluated: Covidien PB 840, Maquet Servo-i, and Dräger Evita XL, PC breaths at preset inspiratory time (TI) 0.6-1.4s. VT error was defined as ventilator displayed VT minus the simulator displayed VT (mL/kg or % of true).
Results:
Relaxation of inspiratory effort caused flow reversal (exhalation) during TI, which led to VT error. For the PB 840, VT error was proportional to TI (maximum -2.0mL/kg, -19%). For the Servo-i, VT error was not related to TI (maximum error -0.2mL/kg or -1.2%). Volume error for Evita XL was not related to TI (maximum error was -0.7mL/kg or -6%).
Conclusions:
Calculation of VT as the integral of flow over the preset inspiratory time rather than the period between zero crossings of flow may result in underestimation of both inhaled and exhaled volumes. The size of VT error can be large enough to potentially affect patient outcomes on some ventilators.
Related Concept Videos
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...
Respiratory Volumes and Capacities