Transmission of violet and blue light through conventional (layered) and bulk cured resin-based composites
J E Harlow1, F A Rueggeberg2, D Labrie3
1Dental Clinical Sciences, Dalhousie University, P.O. BOX 15000, Halifax (NS), B3H 4R2, Canada.
Objectives:
This study measured the transmission of light in the 'violet' (350≤λ≤425nm) and 'blue' (425<λ≤550nm) spectral ranges from a polywave(®) LED curing light through different thicknesses of four commercial, resin-based composites (RBCs).
Material And Methods:
Samples of conventional layered RBCs (Tetric EvoCeram A2, Filtek Supreme Ultra A2B), and bulk-curing resins (Tetric EvoCeram Bulk Fill IVA, and SureFil SDR Flow U) were prepared. Three samples of each RBC were made at thicknesses of 0.1, 0.7, 1, 2, and 4-mm. The uncured RBC specimens were affixed at the entrance aperture of a 6-inch integrating sphere and light-cured once for 20s using a polywave(®) LED curing light (Bluephase G2) on its high power setting. The spectral radiant power transmitted through each RBC in the 'violet' and 'blue' regions was measured using a fiberoptic spectrometer.
Results:
As RBC thickness increased, an exponential attenuation of transmitted light was measured (R(2)>0.98). Attenuation was greater for the 'violet' than for the 'blue' spectral regions. At the light tip, the violet light component represented 15.4% of the light output. After passing through 4-mm of RBC, the violet light represented only between 1.2-3.1% of the transmitted light depending on the RBC. Depending on RBC, approximately 100mW from the Bluephase G2 was transmitted through 0.1-mm of RBC in the 'violet' range, falling at most to 11mW after passing through 2-mm of RBC, and to only 2mW at 4-mm depth.
Conclusions:
Increasing RBC thickness results in an exponential decrease in light transmission. This attenuation is RBC-dependent with shorter wavelengths (violet) attenuated to a greater extent than longer wavelengths (blue).
Clinical Relevance:
Despite the increased translucency of bulk curing RBCs, spectral radiant power shorter than 425nm from a curing light is unlikely to be effective at a depth of 4-mm or more.