Related Experiment Video
Updated: Mar 16, 2026

Atomic Absorbance Spectroscopy to Measure Intracellular Zinc Pools in Mammalian Cells
Published on: May 16, 2019
[Graphite furnace atomic absorption spectrometry for determination of thallium in blood]
Objective:
Colloidal palladium was used as chemical modifier in the determination of blood thallium by graphite furnace atomic absorption spectrometry.
Methods:
Blood samples were precipitated with 5% (V/V)nitric acid, and then determined by GFAAS with colloidal palladium used as a chemical modifier. 0.2% (W/V)sodium chloride was added in the standard series to improve the matrix matching between standard solution and sample.
Results:
The detection limit was 0.2 μg/L. The correlation coefficient was 0.9991. The recoveries were between 93.9% to 101.5%.The relative standard deviations were between 1.8% to 2.7%.The certified reference material of whole blood thallium was determined and the result was within the reference range Conclusion: The method is accurate, simple and sensitive, and it can meet the needs of detection thallium in blood entirely.