Comparison of a new optical biometer using swept-source optical coherence tomography and a biometer using optical
Kenneth J Hoffer1, Peter C Hoffmann1, Giacomo Savini1
1From the Stein Eye Institute, University of California, Los Angeles, and St. Mary's Eye Center (Hoffer), Santa Monica, California, USA; the Augen und Laserklinik (Hoffmann), Castrop-Rouxelle, Germany; the G.B. Bietti Foundation IRCCS (Savini), Rome, Italy.
The IOLMaster 700 (SS-OCT) and Lenstar LS 900 (OLCR) show excellent agreement for ocular measurements. Clinically insignificant differences in axial length and anterior chamber depth do not impact intraocular lens power calculations.
Area of Science:
- Ophthalmology
- Biomedical Optics
- Medical Device Technology
Background:
- Accurate biometry is crucial for refractive outcomes after cataract surgery.
- Comparing different optical technologies for ocular measurements is essential for clinical practice.
- Swept-source optical coherence tomography (SS-OCT) and optical low-coherence reflectometry (OLCR) are common biometry methods.
Purpose of the Study:
- To evaluate the agreement between the IOLMaster 700 (SS-OCT) and Lenstar LS 900 (OLCR) for ocular biometry.
- To assess the clinical significance of any observed differences in measurements.
- To determine the impact of measurement differences on intraocular lens (IOL) power calculations.
Main Methods:
- Prospective evaluation of 183 eyes using both IOLMaster 700 and Lenstar LS 900.
- Measurements included axial length (AL), keratometry (K), anterior chamber depth (ACD), lens thickness, and central corneal thickness (CCT).
- Statistical analyses included paired t-tests, Pearson correlation, and Bland-Altman plots; IOL power was calculated using standard formulas.
Main Results:
- Excellent agreement and correlation were found for AL between SS-OCT and OLCR (23.61 mm vs 23.60 mm).
- Statistically significant but clinically insignificant differences were noted for AL and ACD.
- No statistically significant differences were found for K, lens thickness, or CCT; IOL power calculations showed no significant difference between devices.
Conclusions:
- The IOLMaster 700 and Lenstar LS 900 demonstrate very good agreement for ocular biometry.
- Observed statistically significant differences are clinically insignificant and do not affect IOL power calculations.
- Both SS-OCT and OLCR devices are reliable for pre-operative cataract surgery measurements.