Vertical Misfit of Single-Implant Abutments Made from Different Materials Under Cyclic Loading
Purpose:
Little information is available in the literature regarding the accuracy of marginal fit of abutments made from different materials. The aim of this study was to compare the level of vertical misfit in abutments made from different materials before and after cyclic loading.
Materials And Methods:
Thirty external hexagon implants, 3.75-mm diameter and 13-mm length, were embedded in epoxy resin and divided into three groups according to abutment type (n = 10): (1) machined titanium (Ti) abutments; (2) premachined gold-alloy (Au) abutments; and (3) machined zirconia (Zr) abutments. The abutments were attached to their respective implants by titanium alloy screws at the manufacturer's recommended torque. A cyclic loading (0.5 × 10(6) cycles; 15 Hz) between 11 and 211 N was applied at an angle of 30 degrees to the long axis of the implants. Implant-abutment vertical misfits (μm) were measured at four predetermined points before and after loading with a stereomicroscope at 60× magnification. The group means were compared using a three-way repeated-measure analysis of variance (ANOVA) and Tukey test (α = .05).
Results:
The results from the ANOVA identified significant effects of the abutment type (P < .001). However, there were no significant effects of cyclic loading (P = .978) or measured surfaces (P = .955). When the abutment groups were compared regardless of cyclic loading and measured surfaces, the Au group showed the highest value of vertical misfit (14.93 ± 0.78), followed by the Ti group (8.53 ± 0.44) and the Zr group (5.64 ± 0.73), with statistically significant differences among them (P < .05).
Conclusion:
The abutment groups showed statistically significant differences in vertical misfit. Zr abutments showed the highest accuracy in terms of marginal fit, and Au abutments showed the lowest. Cyclic loading did not significantly change the level of vertical misfit. All experimental groups presented acceptable values of marginal misfit.
More Related Videos
07:09In situ Compressive Loading and Correlative Noninvasive Imaging of the Bone-periodontal Ligament-tooth Fibrous Joint
Published on: March 7, 2014
05:54Real-Time Dynamic Navigation System for the Precise Quad-Zygomatic Implant Placement in a Patient with a Severely Atrophic Maxilla
Published on: October 18, 2021
Related Concept Videos
Normal Strain under Axial Loading
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical...
Eccentric Axial Loading in a Plane of Symmetry
Deformation of Member under Multiple Loadings
In the case of a member with a variable cross-section, the strain is not constant but depends on the position. The deformation of an...
Impact Loading on a Cantilever Beam
When an object is dropped onto the free end of a cantilever, its potential energy due to gravity is...