Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
Dosimetric Verification around High-density Materials for External Beam Radiotherapy.
Makoto Sasaki1, Manabu Nakata, Mitsuhiro Nakamura
1Division of Clinical Radiology Service, Kyoto University Hospital.
The Acuros XB (AXB) algorithm shows improved dose calculation accuracy around high-density materials in radiation treatment planning systems (RTPS), comparable to Monte Carlo simulations, though it underestimates backscatter dose.
Area of Science:
- Medical Physics
- Radiation Oncology
- Computational Dosimetry
Background:
- Radiation treatment planning systems (RTPS) often lack accuracy in dose distribution calculations near high-density materials.
- The Acuros XB (AXB) algorithm, based on the linear Boltzmann transport equation, is a recent advancement for dose calculation.
Purpose of the Study:
- To evaluate the dose calculation accuracy of the AXB algorithm around high-density materials (titanium and stainless steel) using three X-ray energies.
- To compare AXB's accuracy against other RTPS algorithms (AAA, XVMC) and Monte Carlo (MC) simulations using EBT3 measurements.
Main Methods:
- Inserted titanium and stainless steel into a water-equivalent phantom and measured depth dose using EBT3 film.
- Created a virtual phantom in RTPS, calculating dose distributions with AXB, AAA, XVMC, and MC.
- Compared calculated doses with measured values.
Main Results:
- AXB demonstrated dose calculation accuracy comparable to MC and EBT3 measurements on the exit side of high-density materials.
- AXB underestimated dose by up to 30% at the plane of incidence due to limitations in estimating backscatter effects.
Conclusions:
- The Acuros XB algorithm offers improved accuracy for dose calculations near high-density materials in RTPS compared to traditional algorithms.
- Further refinement of AXB may be needed to fully account for backscatter phenomena for enhanced accuracy at the point of incidence.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Related Concept Videos
Biological Effects of Radiation
Radiation: Applications
The average...
Absorption of Radiation