Evaluation of Acoustic Radiation Force Impulse Imaging (ARFI) for the Determination of Liver Stiffness Using
1Klinik für Gastroenterologie, Hepatologie und Infektiologie, Uniklinik Düsseldorf, Düsseldorf, Germany.
Purpose:
Transient elastography (Fibroscan(©); (FS)) and acoustic radiation force impulse imaging (ARFI) represent noninvasive, user-friendly and quick methods providing an objective and reproducible measure of liver stiffness. The aim of the study was to evaluate cut-off values and performance of ARFI measurements in children using transient elastography as a reference.
Methods/Patients:
A total of 198 children were enrolled in this study. All patients underwent liver stiffness measurements with FS (FS-LS) as well as ARFI (with shear wave velocity quantification; ARFI-SWV) and the performance of ARFI in comparison to FS was studied.
Results:
Significantly higher rates of successful measurements were found for ARFI compared to FS (198/198 (100%) vs. 160/198 (80.8%); p<0.001). ARFI-SWV correlated significantly with FS-LS (r=0.751, p=0.001). ARFI-SWV increased significantly with the stage of fibrosis (1.19+0.15 m/s for patients with FS-LS<7.6 kPa); 1.34+0.22 m/s for patients with 7.6