Thyroglobulin Liquid Chromatography-Tandem Mass Spectrometry Has a Low Sensitivity for Detecting Structural Disease
Umal Azmat1, Kyle Porter2, Leigha Senter3
11 Division of Endocrinology, Diabetes, and Metabolism, The Ohio State University and Arthur G. James Cancer Center , Columbus, Ohio.
Background:
Thyroglobulin (Tg) measurement in patients with positive antithyroglobulin antibodies (anti-TgAbs) is not reliable. Tg measurement using liquid chromatography-tandem mass spectrometry (LC/MS) may be useful in this setting.
Methods:
This is a retrospective study with the objective of determining the accuracy of Tg-LC/MS in patients with thyroid cancer with anti-TgAbs. All patients with follicular cell-derived thyroid cancer (TC) who had thyroglobulin measured using LC/MS assay from November 1, 2013, to November 7, 2014, were evaluated. The frequency of detectable Tg-LC/MS was evaluated, with a functional sensitivity (FS) of 0.5 ng/mL in patients with structural disease. Then performance of Tg-LC/MS versus Tg immunometric assay (IMA) was compared using either Immulite assay (Tg-1) with a FS of 0.9 ng/mL or Beckman assay (Tg-B) with a FS of 0.1 ng/mL in detecting structural disease in patients with positive anti-TgAbs.
Results:
A total of 154 consecutive patients were included in this evaluation. Of these, 116 (75%) patients were positive for anti-TgAbs. In patients with structural disease and positive anti-TgAbs, Tg-LC/MS was undetectable in 43.7% of patients. Then the diagnostic accuracy for structural disease of Tg-LC/MS was compared with each Tg-IMA assay separately. In the 26 patients with positive anti-TgAbs where a Tg-I assay was used, the sensitivity and specificity for detecting structural disease were 33.3% and 88.2%, respectively, for the Tg-I assay, and 44.4% and 94.1%, respectively, for the Tg-LC/MS assay. In the 74 patients with positive anti-TgAbs where Tg-B was used, the sensitivity and specificity for detection of structural disease were 72.7% and 71.4%, respectively, for the Tg-B assay, and 62.6% and 93.7%, respectively, for the Tg-LC/MS assay.
Conclusion:
In patients with thyroid cancer with positive anti-TgAbs, Tg-LC/MS was frequently undetectable and was less sensitive for detecting disease than a Tg assay was with a functional sensitivity of 0.1 ng/mL. For patients with detectable Tg-LC/MS and anti-TgAbs, use of the assay for monitoring requires further prospective studies.
Related Concept Videos
Synthesis and Regulation of Thyroid Hormones
Upon reaching the thyroid gland, TSH stimulates the follicular cells' active uptake of iodide ions from the blood. The ions diffuse to the apical surface of the cells and are oxidized to iodine. The...
Myasthenia Gravis: Diagnostic Tests
The edrophonium test is a diagnostic tool for myasthenia gravis. It involves...
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Antibody Structure
Antibodies, also known as immunoglobulins (Ig), are essential players of the adaptive immune system. These antigen-binding proteins are produced by B cells and make up 20 percent of the total blood plasma by weight. In mammals, antibodies fall into five different classes, which each elicits a different biological response upon antigen binding.
The Y-Shaped Structure of Antibodies Consists of Four Polypeptide Chains
Antibodies consist of four polypeptide chains: two identical heavy...