Effects of refractive index mismatch on SRS and CARS microscopy
Abstract:
An inhomogeneous linear refractive index profile, such as that occurring in biological tissues, is shown to significantly alter stimulated Raman scattering (SRS) and coherent anti-Stokes Raman scattering (CARS) microscopy images. Our finite-difference time-domain simulations show that near-field enhancement and microlensing can lead to an increase of an object's perceived molecular density by a factor of nine and changes in its perceived position by 0.4 μm up to 1.0 μm. Thus the assumption that SRS scales linearly and CARS quadratically with density does not always hold. Furthermore, the inhomogeneous linear index can cause false CARS and AM-SRS signals, even for a homogeneous nonlinear susceptibility.
Related Concept Videos
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and...
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...