188Re image performance assessment using small animal multi-pinhole SPECT/PET/CT system
Pedro L Esquinas1, Cristina Rodríguez-Rodríguez2, José Carlos De La Vega3
1Department of Physics and Astronomy, University of British Columbia, Vancouver, BC V6T 1Z1, Canada.
Summary
This study shows that the VECTor/CT system can produce accurate 188Re images comparable to 99mTc. The high-energy ultra-high resolution collimator (HE-UHRC) is recommended for 188Re imaging due to better performance.
Area of Science:
- Nuclear Medicine
- Medical Imaging
- Radiochemistry
Background:
- Pre-clinical SPECT/PET/CT systems are crucial for evaluating novel radiotracers.
- Rhenium-188 (188Re) is a promising radionuclide for targeted radionuclide therapy and imaging.
- Optimizing imaging protocols for 188Re is essential for accurate pre-clinical studies.
Purpose of the Study:
- To evaluate the performance of a pre-clinical SPECT/PET/CT system for 188Re imaging.
- To compare the efficacy of two multi-pinhole collimators: ultra-high resolution collimator (UHRC) and high-energy ultra-high resolution collimator (HE-UHRC).
Main Methods:
- Phantom studies using micro-Jaszczak and custom phantoms to assess spatial resolution, contrast, CNR, and quantification accuracy.
- Comparison of 188Re images with 99mTc (gold standard).
- Analysis of energy spectra and Monte Carlo simulations of the UHRC.
- In vivo validation using a mouse model injected with 188Re-microspheres.
Main Results:
- 188Re imaging achieved spatial resolution comparable to 99mTc.
- HE-UHRC demonstrated superior contrast and CNR for 188Re imaging compared to UHRC.
- Quantitative accuracy for 188Re was within 10% error for both collimators, similar to 99mTc.
- 188Re imaging with UHRC was significantly impacted by down-scattered photons, while HE-UHRC showed better performance.
- Mouse studies revealed similar biodistribution patterns for 188Re-microspheres with both collimators.
Conclusions:
- The VECTor/CT system provides quantitatively accurate 188Re images with quality comparable to 99mTc.
- The HE-UHRC is recommended for 188Re imaging on the VECTor/CT system due to its ability to mitigate high-energy photon penetration.
- Optimized collimator selection is critical for high-quality 188Re SPECT imaging in pre-clinical research.