A Clinical Evaluation of a Next Generation, Non-Invasive, Selective Radiofrequency, Hands-Free, Body-Shaping Device
Objective:
The aim of this study was to compare clinical outcomes of a noninvasive selective radiofrequency (RF) eld device (BTL Vanquish METM, BTL Industries Inc., Boston MA) with its predecessor (VanquishTM, BTL Industries Inc., Boston MA). The BTL Vanquish METM system has been thoroughly redesigned for more efficient, predictable and homogenous energy delivery to the targeted tissue.
Materials And Methods:
In this multi-center study, 36 subjects with BMIs under 30 were randomly assigned to be treated in Group A (BTL Vanquish METM) or Group B (VanquishTM) in order to obtain a side by side comparison of the devices' ef cacies. Each subject re- ceived 4 weekly 45-minute treatments with the device determined by their group assignments. Measurements of subject's abdominal fat were taken prior to the first treatment and again four weeks after finishing the final treatment.
Results:
The primary outcome was abdominal fat thickness reduction as measured by ultrasound one month following each subject's final treatment. Thirty four subjects completed the study. Two patients did not complete their treatments due to the reasons unrelated to the study (one from each group). Subjects in Group A treated with BTL Vanquish METM had an abdominal fat thickness reduction of 4.17 mm, or 29.46%, while subjects in Group B treated with VanquishTM had an abdominal fat thickness reduction of only 2.72 mm, or 15.21%. The 4 weekly treatments with BTL Vanquish METM in Group A produced a 53% higher reduction (4.17 mm vs 2.72 mm) of abdominal fat layer thickness than those in Group B. The standard deviation of ultrasound measurements in Groups A and B were 1.42mm and 2.21mm, respectively. Assuming a homogenous response across the entire treatment area, the volume of fat reduced was calculated by multiplying the average measured reduction in fat layer by the surface area of the treatment applicator (2100 cm2; 325.5 square inches). It was calculated that Group A patients lost an average of 0.876 liter (0.23 liquid gallon) of fat, while Group B patients lost 0.571 liter (0.15 liquid gallon) of fat.
Discussions And Conclusions:
The mean difference between the tested groups was statistically significant proving better outcomes in the Vanquish METM than its predecessor. Furthermore, the reduction in standard deviation of fat reduction measurements in Group A vs Group B is evidence that the Vanquish METM provides more consistent performance. J Drugs Dermatol. 2016;15(12):1557-1561.
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Muscle Stimulation Frequency
Wave summation
At low firing rates, motor neurons induce individual twitch contractions in muscle fibers. These twitches...
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...