Related Experiment Video
Updated: Mar 8, 2026

Accuracy in Dental Medicine, A New Way to Measure Trueness and Precision
Published on: April 29, 2014
Establishment of local diagnostic reference levels for quality control in intraoral radiography
Maki Izawa1, Yasuo Harata1, Noriyoshi Shiba1
1Division of Dental Radiology, Department of Diagnostic and Therapeutic Science, Meikai University School of Dentistry, 1-1 Keyakidai, Sakadoshi, Saitama 350-0283 Japan.
Objective:
To investigate the dosage and imaging conditions for patients undergoing intraoral radiography at Meikai University Hospital and establish assurance and quality control data.
Methods:
Tube voltage, exposure time, and air kinetic energy released per unit mass (air kerma) of three intraoral radiography units were measured. To calculate the patient entrance dose (PED) for each radiograph using Insight film, we extracted data for 1063 patients from their exposure records. The PED was compared with the diagnostic reference level (DRL) from the European Commission and the UK.
Results:
The tube voltage of the three units was maintained at 60 ± 2 kV. Differences in exposure time were less than 1.7 % for all units. The air kerma rates were well maintained within a 4.2 % error. Based on the patient data, there were no significant differences in the mean exposure times for males and females for all anatomical sites. The mean PED ranged from 1.09 ± 0.31 mGy for the mandibular incisors to 2.42 ± 0.33 mGy for the maxillary molars. The mean PED at the mandibular molars using InSight film was 1.59 ± 0.20 mGy, being less than the recommended value based on the DRL for intraoral radiography in the UK.
Conclusions:
We concluded that radiographic conditions at the hospital have been properly maintained. This basic quality control data may assist other dental radiation facilities to reduce patient dosage.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Quality Assurance
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...