Related Experiment Video
Updated: Mar 8, 2026

A Simplified Stepwise Approach to Echo Guidance during Percutaneous Mitral Valve Repair
Published on: October 16, 2021
Mediguide-Assisted Transseptal Puncture without Echocardiographic Guidance
Gabriel Ballesteros1, Pablo Ramos Ardanaz1, Renzo Neglia1
1Cardiology Department, Clínica Universidad de Navarra, Pamplona, Spain.
This study introduces a new transseptal puncture (TSP) technique using the MediGuide™ (MG) system, significantly reducing radiation exposure during catheter ablation for atrial fibrillation. The MG-assisted TSP achieved a 96.7% success rate with minimal fluoroscopy time compared to conventional methods.
Area of Science:
- Cardiology
- Medical Devices
- Minimally Invasive Procedures
Background:
- Transseptal puncture (TSP) is crucial for catheter ablation procedures.
- Traditional TSP relies heavily on fluoroscopy, leading to significant radiation exposure for patients and staff.
- Minimizing radiation while maintaining procedural safety is a key challenge in electrophysiology.
Purpose of the Study:
- To describe and evaluate a novel nonfluoroscopic technique for left atrium transseptal puncture (TSP) using the MediGuide™ (MG) tracking system.
- To assess the feasibility, safety, and efficacy of MG-assisted TSP in reducing radiation exposure during atrial fibrillation ablation.
- To compare the outcomes of MG-assisted TSP with conventional fluoroscopically guided TSP.
Main Methods:
- A prospective study involving 31 consecutive patients undergoing MG-assisted percutaneous catheter ablation.
- Utilized the MediGuide™ system for nonfluoroscopic guidance of transseptal needle and guidewire advancement.
- A control group of 31 matched patients undergoing conventional fluoroscopically guided TSP was included for comparison.
- Left atrium access was confirmed by pressure changes and guidewire advancement.
Main Results:
- Sixty-two MG-assisted TSP attempts were performed with a 96.7% success rate, without conversion to conventional techniques.
- Mean total fluoroscopy time was significantly lower in the MG group (26.65 ± 37.97 seconds) compared to the conventional TSP group (129.13 ± 37.77 seconds) (P < 0.001).
- No major complications were observed in either group during the TSP procedure.
Conclusions:
- The MediGuide™ (MG) system enables a feasible and safe technique for transseptal puncture (TSP) with minimal radiation exposure.
- This nonfluoroscopic approach eliminates the need for additional imaging modalities like intracardiac echocardiography.
- MG-assisted TSP significantly reduces fluoroscopy time, offering a safer alternative for atrial fibrillation ablation procedures.
More Related Videos
07:39Use of a Percutaneous Ventricular Assist Device/Left Atrium to Femoral Artery Bypass System for Cardiogenic Shock
Published on: August 16, 2021
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015