Edge-preserving denoising for intra-operative cone beam CT in endovascular aneurysm repair
Yi Liu1, Miguel Castro1, Mathieu Lederlin2
1INSERM, U1099, Rennes, F-35000, France; Université de Rennes 1, LTSI, Rennes, F-35000, France.
Abstract:
C-arm cone-beam computed tomography (CBCT) acquisition during endovascular aneurysm repair (EVAR) is an emergent technology with more and more applications. It offers real time imaging with a stationary patient and provides 3-D information to achieve guidance of intervention. However, there is growing concern on the overall radiation doses delivered to patients all along the endovascular management due to pre-, intra-, and post-operative X-ray imaging. Manufactures may have their low dose protocols to realize reduction of radiation dose, but CBCT with a low dose protocol has too many artifacts, particularly streak artifacts, and decreased contrast-to-noise ratio (CNR). To reduce noise and artifacts, a penalized weighted least-squares (PWLS) algorithm with an edge-preserving penalty is proposed. The proposed method is evaluated by quantitative parameters including a defined signal-to-noise ratio (SNR), CNR, and modulation transfer function (MTF) on clinical CBCT. Comparisons with PWLS algorithms with isotropic, TV, Huber, anisotropic penalties demonstrate that the proposed edge-preserving penalty performs well not only on edge preservation, but also on streak artifacts suppression, which may be crucial for observing guidewire and stentgraft in EVAR.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018