Related Experiment Video
Updated: Mar 6, 2026

An in vivo Rodent Model of Contraction-induced Injury and Non-invasive Monitoring of Recovery
Published on: May 11, 2011
Miniaturized force/torque sensor for in vivo measurements of tissue characteristics
Abstract:
This paper presents the development of a surgical instrument to measure interaction forces/torques with organic tissue during operation. The focus is on the design progress of the sensor element, consisting of a spoke wheel deformation element with a diameter of 12 mm and eight inhomogeneous doped piezoresistive silicon strain gauges on an integrated full-bridge assembly with an edge length of 500 μm. The silicon chips are contacted to flex-circuits via flip chip and bonded on the substrate with a single component adhesive. A signal processing board with an 18 bit serial A/D converter is integrated into the sensor. The design concept of the handheld surgical sensor device consists of an instrument coupling, the six-axis sensor, a wireless communication interface and battery. The nominal force of the sensing element is 10 N and the nominal torque is 1 N-m in all spatial directions. A first characterization of the force sensor results in a maximal systematic error of 4.92 % and random error of 1.13 %.

