Related Experiment Video
Updated: Mar 6, 2026

Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
Reduction in operator radiation exposure during transradial coronary procedures using a simple lead rectangle.
Azriel B Osherov1, Sharon Bruoha1, Avishag Laish Farkash1
1Division of Cardiology, Barzilai Medical Center, Ashkelon, Israel; Faculty of Health Sciences, Ben Gurion University of the Negev, Beer-Sheva, Israel.
A lead attenuator significantly reduces operator radiation exposure during transradial percutaneous coronary intervention (PCI). This simple addition to standard protection offers substantial protection against scattered radiation, potentially preventing long-term injury.
Area of Science:
- Interventional Cardiology
- Radiation Safety
- Medical Physics
Background:
- Transradial access for percutaneous coronary intervention (PCI) is associated with fewer complications but raises concerns about operator radiation exposure.
- Ionizing radiation poses a cumulative risk of injury to healthcare professionals, particularly interventional cardiologists.
- Effective shielding strategies are crucial for minimizing radiation dose during fluoroscopically guided procedures.
Purpose of the Study:
- To evaluate the efficacy of a lead-attenuator in reducing operator radiation exposure during transradial PCI.
- To assess the impact of a lead-attenuator on scattered radiation levels in the catheterization laboratory.
Main Methods:
- A prospective, observational study involving 52 patients undergoing transradial PCI.
- Patients were divided into two groups: standard operator protection (n=26) and standard protection plus a lead-attenuator (n=26).
- Operator radiation exposure at the thyroid level was measured and compared between the groups.
Main Results:
- The lead-attenuator group showed a significant reduction in operator radiation exposure at the thyroid level (20.2%, p < 0.0001).
- Within the attenuator group, radiation was reduced by 94.5% underneath the lead attenuator compared to above it (p < 0.0001).
- Despite similar fluoroscopy times and examination doses, the attenuator provided substantial additional protection.
Conclusions:
- A simple lead rectangle-attenuator significantly lowers radiation exposure for the primary operator during transradial PCI.
- This protective measure may help mitigate long-term risks of radiation-induced injury for interventional cardiologists.
- The lead attenuator should be considered standard equipment in catheterization laboratories due to its effectiveness in reducing scattered radiation.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
11:38Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014