Related Experiment Video
Updated: Mar 6, 2026

In vivo Structural Assessments of Ocular Disease in Rodent Models using Optical Coherence Tomography
Published on: July 24, 2020
Microscope-Integrated OCT Feasibility and Utility With the EnFocus System in the DISCOVER Study
This study shows that a new microscope-integrated intraoperative optical coherence tomography (OCT) system is feasible for ophthalmic surgery. The system provided valuable insights, changing surgical management in 16% of cases.
Area of Science:
- Ophthalmology
- Medical Imaging
- Surgical Technology
Background:
- Evaluating a novel microscope-integrated intraoperative optical coherence tomography (OCT) system.
- Assessing the feasibility and utility of this technology in ophthalmic surgery.
- Focusing on the EnFocus prototype system within the DISCOVER study.
Purpose of the Study:
- To determine the feasibility and utility of a microscope-integrated intraoperative OCT system.
- To evaluate the impact of intraoperative OCT on surgical decision-making.
- To assess the EnFocus prototype system's performance in ophthalmic procedures.
Main Methods:
- The DISCOVER study involved imaging 50 eyes with the EnFocus intraoperative OCT system.
- OCT imaging was performed at surgeon-determined points during ophthalmic surgery.
- Surgeons completed post-operative questionnaires to report OCT's influence on management.
Main Results:
- Successful imaging was achieved in 92% of cases (46 out of 50 eyes).
- Intraoperative OCT findings led to changes in surgical management in 16% of cases (8 out of 50).
- In membrane peeling, OCT findings differed from initial surgical impressions in 35% of cases (7 out of 20).
Conclusions:
- The study confirms the feasibility of microscope-integrated intraoperative OCT with the Bioptigen EnFocus system.
- Intraoperative OCT can offer surgeons supplementary information, potentially altering surgical decisions.
- This technology holds promise for enhancing ophthalmic surgical procedures.
More Related Videos
06:19In Vivo Imaging of Cx3cr1gfp/gfp Reporter Mice with Spectral-domain Optical Coherence Tomography and Scanning Laser Ophthalmoscopy
Published on: November 11, 2017
12:54Simultaneous Brightfield, Fluorescence, and Optical Coherence Tomographic Imaging of Contracting Cardiac Trabeculae Ex Vivo
Published on: October 2, 2021
Related Concept Videos
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Confocal Fluorescence Microscopy
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Overview of Microscopy Techniques