A candidate reference method for serum potassium measurement by inductively coupled plasma mass spectrometry
Ying Yan1, Bingqing Han1, Jie Zeng1
1.
Background:
Potassium is an important serum ion that is frequently assayed in clinical laboratories. Quality assurance requires reference methods; thus, the establishment of a candidate reference method for serum potassium measurements is important.
Methods:
An inductively coupled plasma mass spectrometry (ICP-MS) method was developed. Serum samples were gravimetrically spiked with an aluminum internal standard, digested with 69% ultrapure nitric acid, and diluted to the required concentration. The 39K/27Al ratios were measured by ICP-MS in hydrogen mode. The method was calibrated using 5% nitric acid matrix calibrators, and the calibration function was established using the bracketing method.
Results:
The correlation coefficients between the measured 39K/27Al ratios and the analyte concentration ratios were >0.9999. The coefficients of variation were 0.40%, 0.68%, and 0.22% for the three serum samples, and the analytical recovery was 99.8%. The accuracy of the measurement was also verified by measuring certified reference materials, SRM909b and SRM956b. Comparison with the ion selective electrode routine method and international inter-laboratory comparisons gave satisfied results.
Conclusions:
The new ICP-MS method is specific, precise, simple, and low-cost, and it may be used as a candidate reference method for standardizing serum potassium measurements.
More Related Videos
10:38Simultaneous Quantification of Selected Kynurenines Analyzed by Liquid Chromatography-Mass Spectrometry in Medium Collected from Cancer Cell Cultures
Published on: May 9, 2020
09:22Absolute Quantitation of Inositol Pyrophosphates by Capillary Electrophoresis Electrospray Ionization Mass Spectrometry
Published on: August 13, 2021
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Atomic Emission Spectroscopy: Lab
Flame Photometry: Lab