Choice of reference measurements affects quantification of long diffusion time behaviour using stimulated echoes
Michiel Kleinnijenhuis1, Jeroen Mollink1,2, Wilfred W Lam1,3
1Oxford Centre for Functional MRI of the Brain, University of Oxford, Oxford, United Kingdom.
Magnetic Resonance in Medicine
|May 5, 2017
Summary
Reference data b-values in diffusion-weighted stimulated echo acquisition mode (DW-STEAM) can create artificial diffusion time dependence in apparent diffusion coefficient (ADC) measurements. A modification is proposed to mitigate this effect in long diffusion time imaging.
Area of Science:
- Magnetic Resonance Imaging (MRI)
- Diffusion Tensor Imaging (DTI)
- Quantitative MRI
Background:
- Diffusion-weighted stimulated echo acquisition mode (DW-STEAM) is used for long diffusion time measurements.
- Accurate quantification of the apparent diffusion coefficient (ADC) is crucial in diffusion MRI.
- Reference data acquisition in DW-STEAM presents challenges due to the inability to achieve a true b=0 value.
Purpose of the Study:
- To investigate the impact of reference data b-values on ADC quantification in long diffusion time DW-STEAM.
- To identify and address the contribution of crusher gradients in DW-STEAM measurements.
- To propose a modification for improved ADC quantification by avoiding crusher gradient artifacts.
Main Methods:
- Two reference data acquisition strategies were evaluated: constant diffusion weighting (fixed-b0) and constant gradient area (fixed-q0).
- Signal calculations were performed for systems with one and two diffusion coefficients.
- Experimental validation was conducted using postmortem human corpus callosum samples.
Main Results:
- Underestimation of ADC was observed for reference images with b>0, leading to apparent time-dependence with fixed-q0.
- Restricted diffusion systems were also affected by this phenomenon.
- Experimental results showed that fixed-q0 exaggerated diffusion time-dependent effects by 62% (longitudinal) and 35% (radial) compared to fixed-b0 in white matter.
Conclusions:
- Variations in reference measurement b-values in DW-STEAM can introduce artificial diffusion time dependence in ADC.
- This artifact occurs even in the absence of diffusion restriction.
- The findings highlight the importance of careful reference data selection for accurate ADC quantification in long diffusion time MRI.
Keywords:
apparent diffusion coefficientbiexponential diffusiondiffusion timerestrictionstimulated echo