A Reproducible Computerized Method for Quantitation of Capillary Density using Nailfold Capillaroscopy
Published on: October 27, 2015
Quantitative outcome measures for systemic sclerosis-related Microangiopathy - Reliability of image acquisition in
Graham Dinsdale1, Tonia Moore2, Neil O'Leary3
1Division of Musculoskeletal & Dermatological Sciences, University of Manchester, Salford Royal Hospital NHS Foundation Trust, Salford Royal Hospital, Stott Lane, Manchester, UK.
Background:
Nailfold capillaroscopic parameters hold increasing promise as outcome measures for clinical trials in systemic sclerosis (SSc). Their inclusion as outcomes would often naturally require capillaroscopy images to be captured at several time points during any one study. Our objective was to assess repeatability of image acquisition (which has been little studied), as well as of measurement.
Method:
41 patients (26 with SSc, 15 with primary Raynaud's phenomenon) and 10 healthy controls returned for repeat high-magnification (300×) videocapillaroscopy mosaic imaging of 10 digits one week after initial imaging (as part of a larger study of reliability). Images were assessed in a random order by an expert blinded observer and 4 outcome measures extracted: (1) overall image grade and then (where possible) distal vessel locations were marked, allowing (2) vessel density (across the whole nailfold) to be calculated (3) apex width measurement and (4) giant vessel count. Intra-rater, intra-visit and intra-rater inter-visit (baseline vs. 1week) reliability were examined in 475 and 392 images respectively. A linear, mixed-effects model was used to estimate variance components, from which intra-class correlation coefficients (ICCs) were determined.
Results:
Intra-visit and inter-visit reliability estimates (ICCs) were (respectively): overall image grade, 0.97 and 0.90; vessel density, 0.92 and 0.65; mean vessel width, 0.91 and 0.79; presence of giant capillary, 0.68 and 0.56. These estimates were conditional on each parameter being measurable.
Conclusion:
Within-operator image analysis and acquisition are reproducible. Quantitative nailfold capillaroscopy, at least with a single observer, provides reliable outcome measures for clinical studies including randomised controlled trials.