Related Experiment Video
Updated: Mar 1, 2026

Angle-resolved Photoemission Spectroscopy At Ultra-low Temperatures
Published on: October 9, 2012
Multipurpose setup for low-temperature conversion electron Mössbauer spectroscopy
V Augustyns1, M Trekels1, H P Gunnlaugsson1
1Instituut voor Kern- en Stralingsfysica, KU Leuven, 3001 Leuven, Belgium.
Abstract:
We describe an experimental setup for conversion electron Mössbauer spectroscopy (CEMS) at low temperature. The setup is composed of a continuous flow cryostat (temperature range of 4.2-500 K), detector housing, three channel electron multipliers, and corresponding electronics. We demonstrate the capabilities of the setup with CEMS measurements performed on a sample consisting of a thin enriched 57Fe film, with a thickness of 20 nm, deposited on a silicon substrate. We also describe exchangeable adaptations (lid and sample holder) which extend the applicability of the setup to emission Mössbauer spectroscopy as well as measurements under an applied magnetic field.