Using Radar Plots to Demonstrate the Accuracy and Precision of 6 Blood Glucose Monitoring Systems
Scott Pardo1, Nancy Dunne1, David A Simmons1
11 Ascensia Diabetes Care, Parsippany, NJ.
Journal of Diabetes Science and Technology
|June 13, 2017
Summary
Radar Plots visually compare blood glucose monitoring system (BGMS) accuracy. The CONTOUR®NEXT (CN) BGMS demonstrated high accuracy, with 99.6% of results meeting ISO 15197:2013 criteria, outperforming other systems evaluated.
Area of Science:
- Medical Devices
- Biomedical Engineering
- Clinical Diagnostics
Background:
- Previous evaluations compared fingertip capillary blood glucose monitoring systems (BGMSs) against a YSI glucose analyzer.
- This study introduces Radar Plots for visualizing BGMS accuracy and precision data.
Purpose of the Study:
- To graphically represent accuracy and precision results of BGMSs using Radar Plots.
- To assess if BGMS measurements meet ISO 15197:2013 accuracy criteria.
Main Methods:
- Radar Plots were used to visualize analytical performance data.
- Concentric circles represent error magnitude; data points indicate deviation from reference values.
- The "15-15 Zone" defines acceptable error limits (±15 mg/dL or ±15%).
Main Results:
- The percentage of results within the "15-15 Zone" varied from 83.6% to 99.8% across six BGMSs.
- The CONTOUR®NEXT (CN) BGMS achieved 99.6% of results within the "15-15 Zone".
Conclusions:
- Radar Plots offer a novel method for visually comparing BGMS analytical performance.
- The tight clustering of CN data points on the Radar Plot indicates superior analytical performance compared to other systems.
Related Concept Videos
Errors occurring during blood pressure monitoring
1.6K
Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
Several factors...
1.6K
Equipments Used To Measure Blood Pressure
3.8K
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
3.8K
Receiver Operating Characteristic Plot
519
A ROC (Receiver Operating Characteristic) plot is a graphical tool used to assess the performance of a binary classification model by illustrating the trade-off between sensitivity (true positive rate) and specificity (false positive rate). By plotting sensitivity against 1 - specificity across various threshold settings, the ROC curve shows how well the model distinguishes between classes, with a curve closer to the top-left corner indicating a more accurate model. The area under the ROC curve...
519
Measurement of Blood Pressure
3.6K
Assessing blood pressure is a standard procedure executed in virtually all medical environments. The method utilized today was established over a hundred years ago by an innovative Russian doctor, Dr. Nikolai Korotkoff. The soft ticking noise, known as Korotkoff sounds, heard while taking blood pressure readings results from turbulent blood flow within the vessels. The apparatus required for this procedure includes a sphygmomanometer, a blood pressure cuff attached to a gauge, and a...
3.6K
Assessing Blood pressure using a doppler ultrasound
2.7K
To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
2.7K
Sites for measuring blood pressure
3.9K
Blood pressure measurement is a fundamental clinical procedure, providing crucial data for assessing cardiovascular health. Among the various sites for this measurement, the brachial and popliteal arteries are predominantly utilized due to their accessibility and the reliability of their readings. This lesson delves into the anatomical significance, methodology, and considerations of measuring blood pressure at these locations.
The Brachial Artery: Primary Site for Blood Pressure Measurement
The Brachial Artery: Primary Site for Blood Pressure Measurement
3.9K