Refined heart failure detection algorithm for improved clinical reliability of OptiVol alerts in CRT-D recipients
Mate Vamos1, Noemi Nyolczas, Zsolt Bari
1University Hospital Frankfurt - Goethe University. vamos.mate@gmail.com.
Background:
The reliability of intrathoracic impedance monitoring for prediction of heart failure (HF) by implantable cardiac devices is controversial. Despite using additional device-based parameters described in the PARTNERS HF study, such as new onset of arrhythmias, abnormal autonomics, low biventricular pacing rate or patient activity level, the predictive power of device diagnostic algorithm is still in doubt. The objective of this study was to compare the device diagnostic algorithm described in the PARTNERS HF study to a newly developed algorithm applying refined diagnostic criteria.
Methods:
Fourty two patients were prospectively enrolled who had been implanted with an intrathoracic impedance and remote monitoring capable implantable cardiac defibrillator with a cardiac resychroniza-tion therapy (CRT-D) device in this observational study. If a remote OptiVolTM alert occurred, patients were checked for presence of HF symptoms. A new algorithm was derived from the original PARTNERS HF criteria, considering more sensitive cut-offs and changes of patterns of the device-based parameters.
Results:
During an average follow-up of 38 months, 722 remote transmissions were received. From the total of 128 transmissions with OptiVol alerts, 32 (25%) corresponded to true HF events. Upon multivariate discriminant analysis, low patient activity, high nocturnal heart rate, and low CRT pacing (< 90%) proved to be independent predictors of true HF events (all p < 0.01). Incorporating these three refined criteria in a new algorithm, the diagnostic yield of OptiVol was improved by increasing specific-ity from 37.5% to 86.5%, positive predictive value from 34.1% to 69.8% and area under the curve from 0.787 to 0.922 (p < 0.01), without a relevant loss in sensitivity (96.9% vs. 93.8%).
Conclusions:
A refined device diagnostic algorithm based on the parameters of low activity level, high nocturnal heart rate, and suboptimal biventricular pacing might improve the clinical reliability of OptiVol alerts.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
06:16Signal Acquisition, Score Interpretation, and Economics of a Non-Invasive Point-of-Care Test for Coronary Artery Disease
Published on: August 9, 2024
Related Concept Videos
Heart Failure IV: Classification and Diagnostic Evaluation
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Cardiomyopathy II: Dilated Cardiomyopathy
Heart Failure VI: Adjunct Therapies
Heart Failure III: Clinical Manifestations
Heart Failure V: Medical Management