Related Experiment Video
Updated: Feb 27, 2026

Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
Radiation reduction of minimally invasive transforaminal lumbar interbody fusion with localisation system in
1Tongji University School of Medicine, 301 Yanchang Road, Shanghai 200072, China.
Aims:
Minimally invasive transforaminal lumbar interbody fusion (MITLIF) has been well validated in overweight and obese patients who are consequently subject to a higher radiation exposure. This prospective multicentre study aimed to investigate the efficacy of a novel lumbar localisation system for MITLIF in overweight patients.
Patients And Methods:
The initial study group consisted of 175 patients. After excluding 49 patients for various reasons, 126 patients were divided into two groups. Those in Group A were treated using the localisation system while those in Group B were treated by conventional means. The primary outcomes were the effective radiation dosage to the surgeon and the exposure time.
Results:
There were 62 patients in Group A and 64 in Group B. The mean effective dosage was 0.0217 mSv (standard deviation (sd) 0.0079) in Group A and 0.0383 mSv (sd 0.0104) in Group B (p < 0.001). The mean fluoroscopy exposure time was 26.42 seconds (sd 5.91) in Group A and 40.67 seconds (sd 8.18) in Group B (p < 0.001). The operating time was 175.56 minutes (sd 32.23) and 206.08 minutes (sd 30.15) (p < 0.001), respectively. The mean pre-operative localisation time was 4.73 minutes (sd 0.84) in Group A and 7.03 minutes (sd 1.51) in Group B (p < 0.001). The mean screw placement time was 47.37 minutes (sd 10.43) in Group A and 67.86 minutes (sd 14.15) in Group B (p < 0.001). The pedicle screw violation rate was 0.35% (one out of 283) in Group A and 2.79% (eight out of 287) in Group B (p = 0.020).
Conclusion:
The study shows that the localisation system can effectively reduce radiation exposure, exposure time, operating time, pre-operative localisation time, and screw placement time in overweight patients undergoing MITLIF. Cite this article: Bone Joint J 2017;99-B:944-50.