Local breast density assessment using reacquired mammographic images
Eloy García1, Oliver Diaz1, Robert Martí1
1Computer Vision and Robotics Institute, University of Girona, Spain.
European Journal of Radiology
|July 3, 2017
Summary
Volpara™ Density Maps reliably assesses glandular tissue distribution in mammograms, showing robustness to minor changes in acquisition angle and beam energy, crucial for follow-up assessments.
Area of Science:
- Radiology and Medical Imaging
- Biomedical Engineering
- Quantitative Imaging
Background:
- Accurate assessment of glandular tissue distribution in mammography is essential for breast cancer screening and risk evaluation.
- Quantitative imaging techniques offer objective measures of breast composition, complementing traditional density assessments.
Purpose of the Study:
- To evaluate the spatial glandular volumetric tissue distribution and density measures provided by Volpara™ Density Maps.
- To assess the reliability of Volpara™ Density Maps using repeated mammogram pairs with slight variations in acquisition parameters.
Main Methods:
- Retrospective analysis of 99 pairs of full-field digital mammograms.
- Utilized Volpara™ Density Maps software to estimate global and local glandular tissue distribution.
- Employed histogram similarity metrics and difference image statistics to evaluate local glandular information.
Main Results:
- High correlations observed for global measures: breast volume (R=0.99), glandular tissue volume (R=0.94), and volumetric breast density (R=0.96).
- Histogram similarity metrics indicated high information sharing between repeated images.
- Small changes in acquisition angle did not significantly alter the glandular pattern, but breast thickness variations did.
Conclusions:
- Volpara™ Density Maps demonstrates reliability in estimating local glandular tissue distribution for assessment and follow-up.
- The software is robust to minor variations in acquisition angle and beam energy.
- Divergences in results can occur due to differing breast compression conditions.