Related Experiment Video
Updated: Feb 27, 2026

In vivo Imaging of Biological Tissues with Combined Two-Photon Fluorescence and Stimulated Raman Scattering Microscopy
Published on: December 20, 2021
Design of spectral-spatial phase prewinding pulses and their use in small-tip fast recovery steady-state imaging
Sydney N Williams1, Jon-Fredrik Nielsen1, Jeffrey A Fessler1
1Department of Biomedical Engineering, University of Michigan, Ann Arbor, Michigan, USA.
Purpose:
Spectrally selective "prewinding" radiofrequency pulses can counteract B0 inhomogeneity in steady-state sequences, but can only prephase a limited range of off-resonance. We propose spectral-spatial small-tip angle prewinding pulses that increase the off-resonance bandwidth that can be successfully prephased by incorporating spatially tailored excitation patterns.
Theory And Methods:
We present a feasibility study to compare spectral and spectral-spatial prewinding pulses. These pulses add a prephasing term to the target magnetization pattern that aims to recover an assigned off-resonance bandwidth at the echo time. For spectral-spatial pulses, the design bandwidth is centered at the off-resonance frequency for each spatial location in a field map. We use these pulses in the small-tip fast recovery steady-state sequence, which is similar to balanced steady-state free precession. We investigate improvement of spectral-spatial pulses over spectral pulses using simulations and small-tip fast recovery scans of a gel phantom and human brain.
Results:
In simulation, spectral-spatial pulses yielded lower normalized root mean squared excitation error than spectral pulses. For both experiments, the spectral-spatial pulse images are also qualitatively better (more uniform, less signal loss) than the spectral pulse images.
Conclusion:
Spectral-spatial prewinding pulses can prephase over a larger range of off-resonance than their purely spectral counterparts. Magn Reson Med 79:1377-1386, 2018. © 2017 International Society for Magnetic Resonance in Medicine.
Related Concept Videos
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...