Establishing national diagnostic reference levels (DRLs) for computed tomography in Egypt
Dina Husseiny Salama1, Jenia Vassileva2, Gamal Mahdaly3
1Egyptian Atomic Energy Authority, Cairo, Egypt.
Egypt established national computed tomography (CT) diagnostic reference levels (DRLs). While CTDIvol is lower than international levels, higher dose-length product (DLP) suggests potential for scan range optimization.
Area of Science:
- Radiology and Imaging
- Medical Physics
- Public Health
Background:
- Diagnostic Reference Levels (DRLs) are crucial for optimizing radiation dose in medical imaging.
- Establishing national DRLs ensures consistent radiation protection standards across healthcare facilities.
Purpose of the Study:
- To determine national DRLs for adult computed tomography (CT) examinations in Egypt.
- To identify opportunities for radiation dose optimization in Egyptian CT facilities.
Main Methods:
- Collected CTDIvol and DLP data from 3762 scans across 50 facilities in Egypt.
- Covered head, chest, abdomen, abdomen-pelvis, chest-abdomen-pelvis, and CT angiography examinations.
- Compared Egyptian DRLs with international benchmarks from the UK, USA, Canada, Japan, Australia, and France.
Main Results:
- Established Egyptian DRLs for CTDIvol (mGy): Head (30), Chest (22), Abdomen (31), Abdomen-Pelvis (31), Chest-Abdomen-Pelvis (33), CT Angiography (37).
- Established Egyptian DRLs for DLP (mGy.cm): Head (1360), Chest (420), Abdomen (1425), Abdomen-Pelvis (1325), Chest-Abdomen-Pelvis (1320), CT Angiography (1320).
- Egyptian head CT CTDIvol DRLs are lower than international levels, but DLP DRLs are comparable or higher.
Conclusions:
- Egyptian DRLs for CTDIvol are generally lower than international standards.
- Higher dose-length product (DLP) DRLs indicate a potential issue with scan range in Egyptian CT protocols.
- Optimization efforts should focus on reducing scan range to align DLP with international benchmarks.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System V: CT